Since line broadening due to plasma processes or the instrument degrade spectral resolution, leading to uncertainty in the elemental profile obtained by optical emission spectroscopy, the current study introduced a novel approach to quantify spectral interferences in laser-induced breakdown spectroscopy (LIBS).
The proposed algorithm establishes a statistical interference factor (SIF) to quantify spectral interferences for individual line assignment in LIBS while using the full information provided by a spectrum. This factor combines fundamentals of plasma emission with a Bayesian analysis of the experimental spectrum, using eventual prior knowledge about the sample and/or the conditions of analysis. Two types of prior knowledge were used to interpret spectra from a pure silicon sample as well as a NIST SRM 610 glass sample and alumina with a controlled Ni contamination at different concentrations. Knowledge of the elemental composition confirmed the existence of spectral lines with high sensitivity to interference, while other lines were more appropriate for analysis due to their consistent SIFs, despite having knowledge of the composition. (publisher abstract modified)
Downloads
Similar Publications
- Increasing Safety in High Need Schools: An Evaluation of Therapeutic Crisis Intervention for Schools
- Estimation of Population-specific Values of Theta for PowerPlex Y23 Profiles
- Further Development of Raman Spectroscopy for Body Fluid Investigation: Forensic Identification, Limit of Detection, and Donor Characterization