The quantity of mitochondrial DNA (mtDNA) template added for amplification and subsequent dye terminator reactions is critical for obtaining quality sequence data. Validation of a human mtDNA real-time quantitative PCR (qPCR) assay demonstrated its high degree of reproducibility and precision as well as an extremely sensitive threshold of detection (0.0001 pg/ìL or approximately six human mtDNA copies/ìL). A study of 35 nonprobative bone and teeth evidence samples revealed that 20 pg of mtDNA template is recommended for successful HV1 and HV2 sequence analysis; however, as little as 0.013 pg can generate a full mtDNA profile when using enhanced amplification reactions. The assay can also detect PCR inhibition and is useful for identifying samples that may benefit from re-purification. Overall, the assay is an excellent method to quantify mtDNA and is useful for determining the best analytical approach for successful sequencing. Abstract published by arrangement with Wiley.