Electrokinetic injection (EI) is the primary method used in forensic laboratories to load amplified PCR product in capillary electrophoresis for short tandem repeat (STR) fragment separation. Results of the current study show that when the volume of ILS is reduced, the average peak height and number of alleles increased, thereby increasing the sensitivity of the detection method. Sizing anomalies were observed; however, they did not adversely affect accuracy and precision. The method developed in this study offers a simple and universal procedure to increase the alleles detected in forensic STR analysis. Reducing the volume of ILS to achieve greater sensitivity is applicable to all STR amplification kits and capillary electrophoresis instruments currently used in forensic DNA analysis. (Publisher abstract modified)
Downloads
Similar Publications
- Training police for procedural justice: An evaluation of officer attitudes, citizen attitudes, and police-citizen interactions
- DNA-Based Identification of Forensically Important Lucilia (Diptera: Calliphoridae) in the Continental United States
- Development of a 3D-based Automated Firearms Evidence Comparison System