In this work, a systematic characterization of paper spray ionization-mass spectrometry (PSI-MS) performance under variable environmental conditions was undertaken on a field-deployable MS system.
Significant efforts have gone into coupling portable mass spectrometric systems and ambient ionization methods, with a culminating goal of field-based, point-of-need screening of harmful chemicals. While fieldable, ambient MS methods are an intriguing combination, ruggedness of these ionization methods in regard to naturally-occurring environmental conditions while operating in-situ, such as ambient temperature, relative humidity (RH) and wind, bears further investigation, as such reports are limited. In the current work, a systematic characterization of paper spray ionization-mass spectrometry (PSI-MS) performance under variable environmental conditions was undertaken on a field-deployable MS system. Of all conditions, the presence of wind speeds greater than 3 m/s (∼7 MPH) was shown to have the most substantial effect on the spectral attributes of signal intensity and duration of the PSI-MS spraying event. Generally, moderate ambient temperature and %RH produced the best spectral performance for PSI-MS. Heat maps were generated from field-collected data across variable temperature and %RH in an effort to inform general PSI-MS performance trends and guide on-site operation by potential users. (Publisher Abstract Provided)
Downloads
Similar Publications
- Estimation of Population-specific Values of Theta for PowerPlex Y23 Profiles
- Utilizing Derivatizing Agents for the Differentiation of Cannabinoid isomers in Complex Food, Beverage and Personal-care Product Matrices by Ambient Ionization Mass Spectrometry
- Development and Evaluation of a Nontargeted Electrochemical Surface-Enhanced Raman Spectroscopy (EC-SERS) Screening Method Applied to Authentic Forensic Seized Drug Casework Samples